Cuma, Mart 29, 2024

Portable Mercury Analyzer

The EMP-3 is lightweight, compact, and highly reliable portable continuous mercury analyzer.

A portable atomic absorption spectrophotometer is installed in the equipment main body, so the equipment is easy to carry to any site, where you can ensure simple and quick measurement of elemental vapor phase mercury in the working environment.

The EMP-3 provides a variety of applications that suggest new mercury measurement methods that can change conventional practice.

The sample gas is directly taken into the absorption cell, and mercury concentration is measured by atomic absorption spectroscopy at absorbance of 253.7 nm wavelength. Both weight and size are reduced furthermore compared to our previous model.

The measurement range is as wide as 0 to 2,000μg/Nm3, and the response time is only 1 second. Measurement data is saved in CSV format in the SD card, which can be easily analyzed. Operation is performed on a touch panel type 4.3 inch color LCD.

By installing the GPS function, you do not need to record the measurement point in the field. In addition, it can be used more reliably with various measurement scenes, such as adopting splashed-proof seamless design or realizing the measurement range twice as wide as the previous models.

The EMP-3 has a function that corrects interference from obstructing substances as the background. Even at sites where the existence of gas that shows positive interference (such as hydrocarbon gas) is suspected, the equipment instantly detects and measures mercury alone and takes up as the data.

By connecting options, it can be used for on-site measurements of samples such as high concentration mercury gas (up to 20,000μg/Nm3) and water. No expensive analytical equipment is needed, and only single unit makes various measurements possible.

Applications

• Fluorescent lamps manufacturing, disposal and recycling.
• Emergency response to mercury spill & cleanup monitoring.
• Mercury contamination source surveying & investigation.
• Industrial/worker safety and hygiene surveying.
• Facility workplace monitoring.
• Hazardous waste inspection and screening.
• Mercury in natural gas screening and measurement.

Technical Specifications

• Measuring principle: Non-dispersive double-beam cold vapor atomic absorption spectrometry.
• Light source: Low-pressure mercury discharge lamp.
• Detector: Semiconductor detector (with a 254 nm band-pass filter).
• Detection limit :0.1 µg/Nm3 (Normal).
• Measuring range: 0–2,000 μg/Nm3; 0–20,000 μg/Nm3 (*) (*) when used with Hi-Lutor option.
• Flow rate during measurement ~ 1 L/min.
• Precision RSD 3% (2 µg/Nm3 or higher) without any option.
• Display: 4.3-inch color LCD.
• Operation: Touchscreen.
• Data output: USB flash drive.
• Alarm method: Alarm buzzer and screen blinking (The alarm setpoint is configurable).
• Other functions Temperature compensation, pressure compensation, GPS information recording.
• Operating conditions: Temperature: 0–45°C, humidity:10–80%RH (No dew condensation).
• Dimensions, weight 207(W)x115(D)x140(H)mm, 1.6kg (including the battery).
• Standard accessories Probe, probe filter (x1), battery (x1), battery charger, shoulder/waist belt, USB flash drive.

 

 

 